| Parent link: | Перспективы развития фундаментальных наук=Prospects of Fundamental Sciences Development: сборник научных трудов XХII Международной конференции студентов, аспирантов и молодых ученых, г. Томск, 22-25 апреля 2025/ Национальный исследовательский Томский политехнический университет ; под ред. И. А. Курзиной [и др.].— .— Томск: Изд-во ТПУ Т. 2 : Химия.— 2025.— С. 160-162 |
|---|---|
| Հիմնական հեղինակ: | |
| Այլ հեղինակներ: | , |
| Ամփոփում: | Заглавие с экрана A gold-modified carbon-containing electrode was developed for the voltammetric determination of povidone-iodine in implant coatings. The study optimized electrochemical detection by evaluating the effects of gold deposition time and voltammetric scan range. Gold was electrochemically deposited using a tetrachloroaurate(III) hydrogen solution, with deposition times of 60, 120, and 200 seconds. The optimal scan range was determined to be 0,6…1,4 V, where the highest sensitivity was achieved due to the influence of gold concentration during its anodic dissolution. Electrodes with a 60-second gold deposition time exhibited superior sensitivity compared to those with longer deposition times.The kinetic profile of povidone-iodine release from implant coatings was studied by immersing samples in a phosphate buffer (pH 6,86) for durations ranging from 30 minutes to 36 hours. Results indicated complete release of povidone-iodine within 10 hours. The proposed voltammetric method offers a reliable approach for quantifying povidone-iodine in implant coatings, providing a valuable tool for quality control in pharmaceutical and biomedical applications, particularly for monitoring antimicrobial drug release Текстовый файл |
| Լեզու: | ռուսերեն |
| Հրապարակվել է: |
2025
|
| Խորագրեր: | |
| Առցանց հասանելիություն: | http://earchive.tpu.ru/handle/11683/132853 |
| Ձևաչափ: | Էլեկտրոնային Գրքի գլուխ |
| KOHA link: | https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=682452 |
MARC
| LEADER | 00000naa2a2200000 4500 | ||
|---|---|---|---|
| 001 | 682452 | ||
| 005 | 20251103141804.0 | ||
| 090 | |a 682452 | ||
| 100 | |a 20251019d2025 k||y0rusy50 ca | ||
| 101 | 0 | |a rus | |
| 102 | |a RU | ||
| 135 | |a drcn ---uucaa | ||
| 200 | 1 | |a Определение повидон-иода в покрытиях имплантатов методом вольтамперометрии |d Determination of povidone-iodine in implant coatings by voltammetry |f Е. И. Левин, Б. А. Лапшина |g науч. рук. Е. В. Дорожко |z eng | |
| 300 | |a Заглавие с экрана | ||
| 320 | |a Список литературы: 5 назв | ||
| 330 | |a A gold-modified carbon-containing electrode was developed for the voltammetric determination of povidone-iodine in implant coatings. The study optimized electrochemical detection by evaluating the effects of gold deposition time and voltammetric scan range. Gold was electrochemically deposited using a tetrachloroaurate(III) hydrogen solution, with deposition times of 60, 120, and 200 seconds. The optimal scan range was determined to be 0,6…1,4 V, where the highest sensitivity was achieved due to the influence of gold concentration during its anodic dissolution. Electrodes with a 60-second gold deposition time exhibited superior sensitivity compared to those with longer deposition times.The kinetic profile of povidone-iodine release from implant coatings was studied by immersing samples in a phosphate buffer (pH 6,86) for durations ranging from 30 minutes to 36 hours. Results indicated complete release of povidone-iodine within 10 hours. The proposed voltammetric method offers a reliable approach for quantifying povidone-iodine in implant coatings, providing a valuable tool for quality control in pharmaceutical and biomedical applications, particularly for monitoring antimicrobial drug release | ||
| 336 | |a Текстовый файл | ||
| 461 | 1 | |0 682243 |t Перспективы развития фундаментальных наук |l Prospects of Fundamental Sciences Development |o сборник научных трудов XХII Международной конференции студентов, аспирантов и молодых ученых, г. Томск, 22-25 апреля 2025 |9 682243 |c Томск |n Изд-во ТПУ |f Национальный исследовательский Томский политехнический университет ; под ред. И. А. Курзиной [и др.] | |
| 463 | 1 | |0 682251 |9 682251 |t Т. 2 : Химия |d 2025 |u conference_tpu-2025-C21_V2.pdf |v С. 160-162 |l Vol. 2 : Chemistry | |
| 610 | 1 | |a труды учёных ТПУ | |
| 610 | 1 | |a электронный ресурс | |
| 610 | 1 | |a povidone-iodine | |
| 610 | 1 | |a voltammetric determination | |
| 610 | 1 | |a gold-modified electrode | |
| 700 | 1 | |a Левин |b Е. И. | |
| 701 | 1 | |a Лапшина |b Б. А. | |
| 702 | 1 | |a Дорожко |b Е. В. |c химик |c инженер Томского политехнического университета |f 1979- |g Елена Владимировна |4 727 | |
| 801 | 0 | |a RU |b 63413507 |c 20251019 |g RCR | |
| 856 | 4 | |z http://earchive.tpu.ru/handle/11683/132853 |u http://earchive.tpu.ru/handle/11683/132853 | |
| 942 | |c CF | ||