A Study of a Method for Determining the Size of a Bremsstrahlung Focal Spot Using a Slit Collimator
| Parent link: | Instruments and Experimental Techniques Vol. 63, iss. 1.— 2020.— [P. 6-12] |
|---|---|
| Main Author: | |
| Corporate Author: | |
| Summary: | Title screen A computer model of a bremsstrahlung source based on a 4-MeV betatron has been corrected upon measuring the focal spot using a new version of the slit-collimator method. Based on this model, an analysis of possible sources of an error in measuring the focal spot is performed, thus providing the reliability of experimental results. Режим доступа: по договору с организацией-держателем ресурса |
| Published: |
2020
|
| Subjects: | |
| Online Access: | https://doi.org/10.1134/S0020441219060204 |