Active thermal feedback for massive cryogenic detectors

Bibliographic Details
Parent link:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment: Scientific Journal
Vol. 444, iss. 1-2.— 2000.— [P. 350-352]
Other Authors: Meier O., Bravin M., Bruckmayer M., Stefano P. Di, Frank T., Loidl M., Meunier P., Probst F., Safran G., Seidel W., Sergeyev I., Sisti M., Stodolsky L., Uchaykin S. V. Sergey Victorovich, Zerle L.
Summary:Title screen
A method to stabilize cryogenic detectors with superconducting phase transition thermometers in their operating point is presented. Measurements of X-ray lines emitted by an 55Fe X-ray fluorescence source showed an improvement in energy resolution from 230 to 133 eV on the 1.5 keV aluminium line with this technique. Furthermore the required set-up allows to simulate real events by injecting heat pulses into the thermometer and in this way to calibrate the detector and to monitor its long-term stability
Режим доступа: по договору с организацией-держателем ресурса
Language:English
Published: 2000
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S016890029901390X
http://elibrary.ru/item.asp?id=242081&
Format: Electronic Book Chapter
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=637206