Spatial Resolution of Images and Effective Focus Size of Bremsstrahlung of an Improved Compact Betatron SEA-7 with Electron Energy of 7 MeV
| Parent link: | Instruments and Experimental Techniques.— .— New York: Springer Science+Business Media LLC Vol. 67, iss. 1.— 2023.— P. 94-100 |
|---|---|
| المؤلف الرئيسي: | |
| مؤلف مشترك: | |
| مؤلفون آخرون: | , |
| الملخص: | Title screen Experimental results of determining the resolution of radiography using the bremsstrahlung from the improved compact SEA-7 betatron with electron energy of 7 MeV are presented. Measurements were made using AGFA NDT D4 PbVacuPac X-ray films and Duplex IQI image quality indicator (model EN 462-5). The obtained results demonstrate a different resolution (approximately 0.16 mm) of thin wire pairs of the indicator in different areas of the radiation cone, i.e., a different horizontal size of the effective radiation focus (approximately 0.2 mm), which must be taken into account when analyzing radiographic images of real objects. The vertical size of the radiation focus (approximately 1.4 mm), which is perpendicular to the plane of the electron orbit, was also estimated Текстовый файл AM_Agreement |
| اللغة: | الإنجليزية |
| منشور في: |
2023
|
| الموضوعات: | |
| الوصول للمادة أونلاين: | https://doi.org/10.1134/S0020441224700246 |
| التنسيق: | الكتروني فصل الكتاب |
| KOHA link: | https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=675359 |
MARC
| LEADER | 00000naa0a2200000 4500 | ||
|---|---|---|---|
| 001 | 675359 | ||
| 005 | 20250403135143.0 | ||
| 090 | |a 675359 | ||
| 100 | |a 20241010d2024 k||y0rusy50 ba | ||
| 101 | 0 | |a eng | |
| 102 | |a US | ||
| 135 | |a drcn ---uucaa | ||
| 181 | 0 | |a i |b e | |
| 182 | 0 | |a b | |
| 183 | 0 | |a cr |2 RDAcarrier | |
| 200 | 1 | |a Spatial Resolution of Images and Effective Focus Size of Bremsstrahlung of an Improved Compact Betatron SEA-7 with Electron Energy of 7 MeV |f M. M. Rychkov, V. V. Kaplin, V. A. Smolyanskiy | |
| 203 | |a Текст |c электронный |b визуальный | ||
| 283 | |a online_resource |2 RDAcarrier | ||
| 300 | |a Title screen | ||
| 320 | |a References: 8 tit | ||
| 330 | |a Experimental results of determining the resolution of radiography using the bremsstrahlung from the improved compact SEA-7 betatron with electron energy of 7 MeV are presented. Measurements were made using AGFA NDT D4 PbVacuPac X-ray films and Duplex IQI image quality indicator (model EN 462-5). The obtained results demonstrate a different resolution (approximately 0.16 mm) of thin wire pairs of the indicator in different areas of the radiation cone, i.e., a different horizontal size of the effective radiation focus (approximately 0.2 mm), which must be taken into account when analyzing radiographic images of real objects. The vertical size of the radiation focus (approximately 1.4 mm), which is perpendicular to the plane of the electron orbit, was also estimated | ||
| 336 | |a Текстовый файл | ||
| 371 | 0 | |a AM_Agreement | |
| 461 | 1 | |t Instruments and Experimental Techniques |c New York |n Springer Science+Business Media LLC | |
| 463 | 1 | |t Vol. 67, iss. 1 |v P. 94-100 |d 2023 | |
| 610 | 1 | |a электронный ресурс | |
| 610 | 1 | |a труды учёных ТПУ | |
| 610 | 1 | |a radiography | |
| 610 | 1 | |a determining the resolution | |
| 700 | 1 | |a Rychkov |b M. M. |c physicist |c Head of the laboratory of Tomsk Polytechnic University, Candidate of technical sciences |f 1977- |g Maksim Mikhailovich |9 16251 | |
| 701 | 1 | |a Kaplin |b V. V. |c physicist |c senior research fellow at Tomsk Polytechnic University |f 1947- |g Valery Viktorovich |9 15693 | |
| 701 | 1 | |a Smolyanskiy |b V. A. |c Specialist in the field of instrument engineering |c Researcher, Tomsk Polytechnic University, Candidate of Technical Sciences |f 1991- |g Vladimir Aleksandrovich |9 20730 | |
| 712 | 0 | 2 | |a National Research Tomsk Polytechnic University |c (2009- ) |9 27197 |4 570 |
| 801 | 0 | |a RU |b 63413507 |c 20241010 |g RCR | |
| 856 | 4 | |u https://doi.org/10.1134/S0020441224700246 |z https://doi.org/10.1134/S0020441224700246 | |
| 942 | |c CR | ||