The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study

Bibliographic Details
Parent link:Sensors
Vol. 22, iss. 5.— 2022.— [2032, 16 p.]
Corporate Author: Национальный исследовательский Томский политехнический университет Школа базовой инженерной подготовки Отделение физической культуры
Other Authors: Rogers B. Bruce, Schaffarczyk M. Marcelle, Clauss M. Martina, Mourot L. Laurent, Gronwald Th. Thomas
Summary:Title screen
The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG. A heterogeneous group of 21 participants performed an incremental cycling ramp to failure with measurements of HRV, before (PRE), during (EX), and after (POST). Results showed excellent correlations between devices for linear indexes with Pearson’s r between 0.98 to 1.0 for meanRR, SDNN, RMSSD, and 0.95 to 0.97 for the non-linear index DFA a1 during PRE, EX, and POST. There was no significant difference in device specific meanRR during PRE and POST. Bland–Altman analysis showed high agreement between devices (PRE and POST: meanRR bias of 0.0 and 0.4 ms, LOA of 1.9 to ?1.8 ms and 2.3 to ?1.5; EX: meanRR bias of 11.2 to 6.0 ms; LOA of 29.8 to ?7.4 ms during low intensity exercise and 8.5 to 3.5 ms during high intensity exercise). The Movesense Medical device can be used in lieu of a reference ECG for the calculation of HRV with the potential to differentiate noise from atrial fibrillation and represents a significant advance in both a HR and HRV recording device in a chest belt form factor for lab-based or remote field-application.
Language:English
Published: 2022
Subjects:
Online Access:https://doi.org/10.3390/s22052032
Format: Electronic Book Chapter
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=668113

MARC

LEADER 00000naa0a2200000 4500
001 668113
005 20250324110144.0
035 |a (RuTPU)RU\TPU\network\39337 
035 |a RU\TPU\network\39301 
090 |a 668113 
100 |a 20220621d2022 k||y0rusy50 ba 
101 0 |a eng 
102 |a CH 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study  |f B. Rogers, M. Schaffarczyk, M. Clauss [et al.] 
203 |a Text  |c electronic 
300 |a Title screen 
320 |a [References: 74 tit.] 
330 |a The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG. A heterogeneous group of 21 participants performed an incremental cycling ramp to failure with measurements of HRV, before (PRE), during (EX), and after (POST). Results showed excellent correlations between devices for linear indexes with Pearson’s r between 0.98 to 1.0 for meanRR, SDNN, RMSSD, and 0.95 to 0.97 for the non-linear index DFA a1 during PRE, EX, and POST. There was no significant difference in device specific meanRR during PRE and POST. Bland–Altman analysis showed high agreement between devices (PRE and POST: meanRR bias of 0.0 and 0.4 ms, LOA of 1.9 to ?1.8 ms and 2.3 to ?1.5; EX: meanRR bias of 11.2 to 6.0 ms; LOA of 29.8 to ?7.4 ms during low intensity exercise and 8.5 to 3.5 ms during high intensity exercise). The Movesense Medical device can be used in lieu of a reference ECG for the calculation of HRV with the potential to differentiate noise from atrial fibrillation and represents a significant advance in both a HR and HRV recording device in a chest belt form factor for lab-based or remote field-application. 
461 |t Sensors 
463 |t Vol. 22, iss. 5  |v [2032, 16 p.]  |d 2022 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
610 1 |a heart rate variability 
610 1 |a wearable 
610 1 |a medical device 
610 1 |a health care 
610 1 |a exercise 
610 1 |a сердечные сокращения 
610 1 |a медицинские приборы 
610 1 |a здравоохранение 
610 1 |a упражнения 
701 1 |a Rogers  |b B.  |g Bruce 
701 1 |a Schaffarczyk  |b M.  |g Marcelle 
701 1 |a Clauss  |b M.  |g Martina 
701 1 |a Mourot  |b L.  |c specialist in the field of physical training and sports  |c Senior Researcher of Tomsk Polytechnic University, Candidate of philological sciences  |f 1977-  |g Laurent  |3 (RuTPU)RU\TPU\pers\41001 
701 1 |a Gronwald  |b Th.  |g Thomas 
712 0 2 |a Национальный исследовательский Томский политехнический университет  |b Школа базовой инженерной подготовки  |b Отделение физической культуры  |3 (RuTPU)RU\TPU\col\23545 
801 0 |a RU  |b 63413507  |c 20220624  |g RCR 
856 4 |u https://doi.org/10.3390/s22052032 
942 |c CF