Evaluating the dependency of neutron spectra and absorbed dose rates on the collimation field size in fast neutron therapy

מידע ביבליוגרפי
Parent link:Heliyon
Vol. 7, iss. 11.— 2021.— [e08274, 5 p.]
מחבר ראשי: Shehada A. Abdullah
מחבר תאגידי: Национальный исследовательский Томский политехнический университет Инженерная школа ядерных технологий Научно-образовательный центр Б. П. Вейнберга
מחברים אחרים: Krivobokov V. P. Valery Pavlovich, Golovkov V. M. Vladimir Mikhailovich
סיכום:Title screen
The aim of this research was to investigate the relationship between the collimator aperture and fast-neutron flux, neutron-energy spectrum and absorbed dose rate. For remote therapy, rather large fluxes of fast neutrons are needed which can create dose levels in the tissues of at least 0.1 Gy/min with a source-patient distance of 1 m. Advantageously for these purposes, the 9Be(d, n) reaction was investigated with deuteron energy of 13.6 MeV. The mean energy of the outgoing neutrons was obtained using the code PACE 4 (LISE++) which gave the value of about 5.2 MeV. The maximum neutron flux was at an energy of about 2.5 MeV. Samples activation analysis was deployed to measure the neutron flux in the energy-region [0-14 MeV]. The experimental works were carried out using Al, Fe, Cu and Cd foils which installed on the collimator apertures. To investigate the neutron spectrum, fluxes, and dose rates absorbed at the position of patients, experiments were conducted for four different neutron irradiation-field sizes, which can be modified by the removable-polyethylene parts. Simulation results obtained by the code MCNP-4C and PACE4 (LISE++) were comparable with the experimental data to some extent with consideration of some uncertainties of PACE4 results. It can be concluded that the neutron flux is depended on the irradiation-field size where the neutron flux output for bigger aperture size was about +25% comparing with the smaller ones. These results could play a significant role in improving the neutron flux and optimizing the collimation system utilized in fast neutron therapy. In addition, this can lead to optimization of irradiation canals installed in the nuclear reactors which employed for production of medical isotopes, material testing and many other applications.
יצא לאור: 2021
נושאים:
גישה מקוונת:https://doi.org/10.1016/j.heliyon.2021.e08274
פורמט: אלקטרוני Book Chapter
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=666376

MARC

LEADER 00000naa0a2200000 4500
001 666376
005 20250215152430.0
035 |a (RuTPU)RU\TPU\network\37580 
090 |a 666376 
100 |a 20211220d2021 k||y0rusy50 ba 
101 0 |a eng 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a Evaluating the dependency of neutron spectra and absorbed dose rates on the collimation field size in fast neutron therapy  |f A. Shehada, V. P. Krivobokov, V. M. Golovkov 
203 |a Text  |c electronic 
300 |a Title screen 
320 |a [References: 17 tit.] 
330 |a The aim of this research was to investigate the relationship between the collimator aperture and fast-neutron flux, neutron-energy spectrum and absorbed dose rate. For remote therapy, rather large fluxes of fast neutrons are needed which can create dose levels in the tissues of at least 0.1 Gy/min with a source-patient distance of 1 m. Advantageously for these purposes, the 9Be(d, n) reaction was investigated with deuteron energy of 13.6 MeV. The mean energy of the outgoing neutrons was obtained using the code PACE 4 (LISE++) which gave the value of about 5.2 MeV. The maximum neutron flux was at an energy of about 2.5 MeV. Samples activation analysis was deployed to measure the neutron flux in the energy-region [0-14 MeV]. The experimental works were carried out using Al, Fe, Cu and Cd foils which installed on the collimator apertures. To investigate the neutron spectrum, fluxes, and dose rates absorbed at the position of patients, experiments were conducted for four different neutron irradiation-field sizes, which can be modified by the removable-polyethylene parts. Simulation results obtained by the code MCNP-4C and PACE4 (LISE++) were comparable with the experimental data to some extent with consideration of some uncertainties of PACE4 results. It can be concluded that the neutron flux is depended on the irradiation-field size where the neutron flux output for bigger aperture size was about +25% comparing with the smaller ones. These results could play a significant role in improving the neutron flux and optimizing the collimation system utilized in fast neutron therapy. In addition, this can lead to optimization of irradiation canals installed in the nuclear reactors which employed for production of medical isotopes, material testing and many other applications. 
461 |t Heliyon 
463 |t Vol. 7, iss. 11  |v [e08274, 5 p.]  |d 2021 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
610 1 |a fast neutrons 
610 1 |a collimation system 
610 1 |a neutron therapy 
610 1 |a absorbed dose rate 
610 1 |a MCNP simulations 
700 1 |a Shehada  |b A.  |g Abdullah 
701 1 |a Krivobokov  |b V. P.  |c Russian physicist  |c professor of Tomsk Polytechnic University (TPU), Doctor of Physical and Mathematical Sciences (DSc)  |f 1948-  |g Valery Pavlovich  |3 (RuTPU)RU\TPU\pers\30416  |9 14757 
701 1 |a Golovkov  |b V. M.  |c physicist  |c Head of the laboratory of Tomsk Polytechnic University, Candidate of physical and mathematical sciences  |f 1950-  |g Vladimir Mikhailovich  |3 (RuTPU)RU\TPU\pers\31871 
712 0 2 |a Национальный исследовательский Томский политехнический университет  |b Инженерная школа ядерных технологий  |b Научно-образовательный центр Б. П. Вейнберга  |3 (RuTPU)RU\TPU\col\23561 
801 2 |a RU  |b 63413507  |c 20211220  |g RCR 
856 4 |u https://doi.org/10.1016/j.heliyon.2021.e08274 
942 |c CF