Using a plane wave approximation in simulation of radiation distribution in an X-ray Talbot interferometer; Bulletin of the Lebedev Physics Institute; Vol. 44, iss. 12
| Parent link: | Bulletin of the Lebedev Physics Institute Vol. 44, iss. 12.— 2017.— [P. 350-352] |
|---|---|
| Údair chorparáideacha: | , |
| Rannpháirtithe: | , , , , , |
| Achoimre: | Title screen This study is devoted to the analysis of the plane-wave approximation applicability to X-ray radiation incident on an object. Based on simple calculations, it is shown that an X-ray tube focal spot 0.4•0.8 mm in size at distances of ~1 m can be considered as a point source; however, the plane-wave approximation for such X-ray radiation with energy E=22keV is valid for radiation source-object distances much longer than 10 m. Режим доступа: по договору с организацией-держателем ресурса |
| Teanga: | Béarla |
| Foilsithe / Cruthaithe: |
2017
|
| Ábhair: | |
| Rochtain ar líne: | https://doi.org/10.3103/S1068335617120028 |
| Formáid: | Leictreonach Caibidil leabhair |
| KOHA link: | https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=665827 |
| Achoimre: | Title screen This study is devoted to the analysis of the plane-wave approximation applicability to X-ray radiation incident on an object. Based on simple calculations, it is shown that an X-ray tube focal spot 0.4•0.8 mm in size at distances of ~1 m can be considered as a point source; however, the plane-wave approximation for such X-ray radiation with energy E=22keV is valid for radiation source-object distances much longer than 10 m. Режим доступа: по договору с организацией-держателем ресурса |
|---|---|