A Portable Device for Rapid Assessment of the Functional State of the Hemostasis System

Bibliographische Detailangaben
Parent link:Biomedical Engineering
Vol. 54, iss. 5.— 2021.— [P. 345–349]
1. Verfasser: Zhukov E. L. Egor Leonidovich
Körperschaft: Национальный исследовательский Томский политехнический университет Исследовательская школа химических и биомедицинских технологий
Weitere Verfasser: Gubarev F. A. Fedor Aleksandrovich, Tyutrin I. I. Ivan Illarionovich
Zusammenfassung:Title screen
We present here the results obtained during the development of a portable device for rapid assessment of the functional state of the homeostasis system based on analysis of capillary blood. Vascular-thrombocytic hemostasis was assessed in volunteers in real time using the device. The results of piezocoagulography conducted using the portable device were compared to the results obtained by low-frequency piezothromboelastography (Mednord ARP-01M). These data showed a strong correlation, evidencing the potential of the method and device developed here for studies of the functional state of the hemostasis system at the patient’s bedside or outside hospital.
Режим доступа: по договору с организацией-держателем ресурса
Sprache:Englisch
Veröffentlicht: 2021
Schlagworte:
Online-Zugang:https://doi.org/10.1007/s10527-021-10036-8
Format: Elektronisch Buchkapitel
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=663786

Ähnliche Einträge