High-resolution phonocardiogram parameters; IOP Conference Series: Materials Science and Engineering; Vol. 1019 : 14th International Forum on Strategic Technology (IFOST 2019)

Bibliographische Detailangaben
Parent link:IOP Conference Series: Materials Science and Engineering
Vol. 1019 : 14th International Forum on Strategic Technology (IFOST 2019).— 2021.— [012025, 7 p.]
Körperschaft: Национальный исследовательский Томский политехнический университет Инженерная школа неразрушающего контроля и безопасности Отделение электронной инженерии
Weitere Verfasser: Zemlyakov I. Yu., Zdanov D. S., Kosteley Y. V., Golobokova E. V., Dikman E. Yu. Ekaterina Yurievna, Bureev A. Sh.
Zusammenfassung:Title screen
The article describes the results of studying and analyzing phonocardiograms (PCGs) obtained during a physiological experiment with Blu-ray standard equipment. It provides the findings of a spectral and spectral-time analysis for signals with a sampling frequency of 10, 44.1 and 192 kHz. It shows that the differences in the PCG spectra of identical signals are unreliable. The article specifies the onset and disappearance moments of the harmonic components of heart sounds. It also provides recommendations on the sampling frequency and bit resolution of digitized PCG signals for telemetric systems.
Sprache:Englisch
Veröffentlicht: 2021
Schlagworte:
Online-Zugang:http://earchive.tpu.ru/handle/11683/64547
https://doi.org/10.1088/1757-899X/1019/1/012025
Format: MixedMaterials Elektronisch Buchkapitel
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=663534

MARC

LEADER 00000naa2a2200000 4500
001 663534
005 20251216135839.0
035 |a (RuTPU)RU\TPU\network\34704 
035 |a RU\TPU\network\34696 
090 |a 663534 
100 |a 20210218d2021 k||y0rusy50 ba 
101 0 |a eng 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a High-resolution phonocardiogram parameters  |f I. Yu. Zemlyakov, D. S. Zdanov, Y. V. Kosteley [et al.] 
203 |a Text  |c electronic 
300 |a Title screen 
320 |a [References: 13 tit.] 
330 |a The article describes the results of studying and analyzing phonocardiograms (PCGs) obtained during a physiological experiment with Blu-ray standard equipment. It provides the findings of a spectral and spectral-time analysis for signals with a sampling frequency of 10, 44.1 and 192 kHz. It shows that the differences in the PCG spectra of identical signals are unreliable. The article specifies the onset and disappearance moments of the harmonic components of heart sounds. It also provides recommendations on the sampling frequency and bit resolution of digitized PCG signals for telemetric systems. 
461 1 |0 (RuTPU)RU\TPU\network\2008  |t IOP Conference Series: Materials Science and Engineering 
463 0 |0 (RuTPU)RU\TPU\network\34482  |t Vol. 1019 : 14th International Forum on Strategic Technology (IFOST 2019)  |o October 14-17, 2019, Tomsk, Russia  |f National Research Tomsk Polytechnic University (TPU) ; eds. R. V. Ostvald, D. Gvozdeva, A. S. Ryabikina  |v [012025, 7 p.]  |d 2021 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
610 1 |a спектральный анализ 
610 1 |a сигналы 
610 1 |a спектры 
610 1 |a телеметрические системы 
701 1 |a Zemlyakov  |b I. Yu. 
701 1 |a Zdanov  |b D. S. 
701 1 |a Kosteley  |b Y. V. 
701 1 |a Golobokova  |b E. V. 
701 1 |a Dikman  |b E. Yu.  |c specialist in the field of electronics  |c Associate Professor of Tomsk Polytechnic University, Candidate of technical sciences  |f 1983-  |g Ekaterina Yurievna  |3 (RuTPU)RU\TPU\pers\45867 
701 1 |a Bureev  |b A. Sh. 
712 0 2 |a Национальный исследовательский Томский политехнический университет  |b Инженерная школа неразрушающего контроля и безопасности  |b Отделение электронной инженерии  |3 (RuTPU)RU\TPU\col\23507 
801 2 |a RU  |b 63413507  |c 20210301  |g RCR 
856 4 |u http://earchive.tpu.ru/handle/11683/64547 
856 4 |u https://doi.org/10.1088/1757-899X/1019/1/012025 
942 |c CF