Riassunto:Title screen
The process of blood plasma coagulation was studied using the laser speckle correlation technique. Coagulation times measured in samples of different volumes using the Minilab 701 commercial coagulometer and the laser speckle correlation technique were compared. It was shown that the laser speckle correlation technique provides comparable accuracy of measurement, while allowing the sample volume to be considerably reduced.
Режим доступа: по договору с организацией-держателем ресурса
Pubblicazione: 2018
Soggetti:
Accesso online:https://doi.org/10.1007/s10527-018-9808-x
Natura: Elettronico Capitolo di libro
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=658506
Descrizione
Riassunto:Title screen
The process of blood plasma coagulation was studied using the laser speckle correlation technique. Coagulation times measured in samples of different volumes using the Minilab 701 commercial coagulometer and the laser speckle correlation technique were compared. It was shown that the laser speckle correlation technique provides comparable accuracy of measurement, while allowing the sample volume to be considerably reduced.
Режим доступа: по договору с организацией-держателем ресурса
DOI:10.1007/s10527-018-9808-x