Hardware and Software Complex for Researching Medical Nanoelectrodes; Advances in Computer Science Research; Vol. 51 : Information Technologies in Science, Management, Social Sphere and Medicine (ITSMSSM 2016)

Библиографические подробности
Источник:Advances in Computer Science Research
Vol. 51 : Information Technologies in Science, Management, Social Sphere and Medicine (ITSMSSM 2016).— 2016.— [P. 130-133]
Главный автор: Kim V. L. Valery Lvovich
Автор-организация: Национальный исследовательский Томский политехнический университет Инженерная школа информационных технологий и робототехники Отделение информационных технологий
Другие авторы: Andreev S. A. Semen Alekseevich
Примечания:Title screen
In medical devices required in electrocardiography, electroencephalography, electromyography, electrooculography, etc. the most important element is the medical electrode intended for measuring biopotentials. The optimal selection of electrodes is possible only after testing them to confirm the claimed manufacturers' parameters or determine the real metrological characteristics. Technical characteristics of measuring instruments produced by domestic producers or supplied by foreign companies are regulated by the State Standard GOST 25995-83. Thus, in accordance with this standard, calibration equipment has to have inherent noise level (basic parameter) up to 15 V in a frequency band up to 10 kHz. However, this requirement does not satisfy modern trends of medical instrument engineering in connection with the appearance of new medical nanoelectrodes on the market, including domestic production. A hardware and software complex for researching medical electrodes (HSCRME) was developed at Tomsk Polytechnic University. It allows measuring the parameters of electrode cells: an electrode potential difference; a drift of the electrode potential difference; a noise voltage; an electromechanical noise; an impedance of electrodes; a polarization voltage. The HSCRME includes a measurement analog channel (AMC) and a digital subsystem being a personal computer. The AMC is configured automatically when the user selects the appropriate mode depending on the measured parameter of the electrode. Cross-platform application framework Qt is the software foundation of the HSCRME. Potential possibilities of the HSCRME are determined by the noise of AMC. The noise applied to the input of AMC did not exceed 1 V (rms) in the frequency range 0...10 kHz. Verification of the complex was carried out by the inductive voltage divider, which is part of State Standard GAT193-2011. This State Standard is stored in VNIIFTRI (Moscow).
Язык:английский
Опубликовано: 2016
Предметы:
Online-ссылка:http://dx.doi.org/10.2991/itsmssm-16.2016.70
Формат: Электронный ресурс Статья
Запись в KOHA:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=657512

MARC

LEADER 00000nla2a2200000 4500
001 657512
005 20240119174805.0
035 |a (RuTPU)RU\TPU\network\24042 
035 |a RU\TPU\network\24040 
090 |a 657512 
100 |a 20180212a2016 k y0engy50 ba 
101 0 |a eng 
102 |a FR 
105 |a y z 100zy 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a Hardware and Software Complex for Researching Medical Nanoelectrodes  |f V. L. Kim, S. A. Andreev 
203 |a Text  |c electronic 
300 |a Title screen 
320 |a [References: p. 133 (10 tit.)] 
330 |a In medical devices required in electrocardiography, electroencephalography, electromyography, electrooculography, etc. the most important element is the medical electrode intended for measuring biopotentials. The optimal selection of electrodes is possible only after testing them to confirm the claimed manufacturers' parameters or determine the real metrological characteristics. Technical characteristics of measuring instruments produced by domestic producers or supplied by foreign companies are regulated by the State Standard GOST 25995-83. Thus, in accordance with this standard, calibration equipment has to have inherent noise level (basic parameter) up to 15 V in a frequency band up to 10 kHz. However, this requirement does not satisfy modern trends of medical instrument engineering in connection with the appearance of new medical nanoelectrodes on the market, including domestic production. A hardware and software complex for researching medical electrodes (HSCRME) was developed at Tomsk Polytechnic University. It allows measuring the parameters of electrode cells: an electrode potential difference; a drift of the electrode potential difference; a noise voltage; an electromechanical noise; an impedance of electrodes; a polarization voltage. The HSCRME includes a measurement analog channel (AMC) and a digital subsystem being a personal computer. The AMC is configured automatically when the user selects the appropriate mode depending on the measured parameter of the electrode. Cross-platform application framework Qt is the software foundation of the HSCRME. Potential possibilities of the HSCRME are determined by the noise of AMC. The noise applied to the input of AMC did not exceed 1 V (rms) in the frequency range 0...10 kHz. Verification of the complex was carried out by the inductive voltage divider, which is part of State Standard GAT193-2011. This State Standard is stored in VNIIFTRI (Moscow). 
461 1 |0 (RuTPU)RU\TPU\network\18167  |t Advances in Computer Science Research 
463 0 |0 (RuTPU)RU\TPU\network\18169  |t Vol. 51 : Information Technologies in Science, Management, Social Sphere and Medicine (ITSMSSM 2016)  |o III International Scientific Conference, May 23-26, 2016, Tomsk, Russia  |o [proceedings]  |f National Research Tomsk Polytechnic University (TPU) ; eds. O. G. Berestneva, A. Tikhomirov, A. Trufanov  |v [P. 130-133]  |d 2016 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
610 1 |a hardware and software complex 
610 1 |a medical electrode 
610 1 |a metrological characteristics 
610 1 |a a noise voltage 
610 1 |a аппаратно-программные комплексы 
610 1 |a медицинские электроды 
610 1 |a метрологические характеристики 
610 1 |a биопотенциалы 
610 1 |a шумы 
700 1 |a Kim  |b V. L.  |c specialist in the field of Informatics and computer engineering  |c Professor of Tomsk Polytechnic University, doctor of technical Sciences  |f 1950-  |g Valery Lvovich  |3 (RuTPU)RU\TPU\pers\34851  |9 18187 
701 1 |a Andreev  |b S. A.  |c Specialist in the field of informatics and computer technology  |c Assistant of the Department of Tomsk Polytechnic University  |f 1993-  |g Semen Alekseevich  |3 (RuTPU)RU\TPU\pers\38420 
712 0 2 |a Национальный исследовательский Томский политехнический университет  |b Инженерная школа информационных технологий и робототехники  |b Отделение информационных технологий  |3 (RuTPU)RU\TPU\col\23515 
801 2 |a RU  |b 63413507  |c 20180212  |g RCR 
856 4 |u http://dx.doi.org/10.2991/itsmssm-16.2016.70 
942 |c CF