Measurement of the current transfer ratio for current shunts; Measurement in Research and Industry

Dettagli Bibliografici
Parent link:Measurement in Research and Industry.— 2015.— [4 р.]
Ente Autore: Томский политехнический университет
Altri autori: Baranov P. F. Pavel Fedorovich, Tsimbalist E. I. Edvard Ilyich, Borikov V. N. Valery Nikolaevich, Soltanova D. G. Diana G.
Riassunto:Title screen
Current shunts are widely used in metrological practice. The shunt is required to determine the value of its resistance on DC and AC currents to determine the transfer ratio error in the wide frequency band in process calibration. The paper describes a procedure for measuring the complex impedance of transfer ratio of current shunts in the frequency band using the lock-in amplifier with the differential input. The experimental results of the proposed procedure are described.
Lingua:inglese
Pubblicazione: 2015
Soggetti:
Accesso online:http://www.imeko.org/publications/wc-2015/IMEKO-WC-2015-TC4-102.pdf
Natura: Elettronico Capitolo di libro
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=656291