Method for studying the linearity of photomultipliers using noncalibrated light sources; Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment; Vol. 875
| Parent link: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment Vol. 875.— 2017.— [P. 137-140] |
|---|---|
| Kurumsal yazarlar: | , |
| Diğer Yazarlar: | , , , |
| Özet: | Title screen A simple method for measuring and evaluation of nonlinearity of the photomultiplier response efficiency using two controlled noncalibrated light sources is presented. The accuracy of the response efficiency nonlinearity evaluation and its linearization procedure was found to be 0.15%. As was shown, photomultiplier nonlinearity led to nonproportional distortions of photoelectronic estimates. Режим доступа: по договору с организацией-держателем ресурса |
| Dil: | İngilizce |
| Baskı/Yayın Bilgisi: |
2017
|
| Konular: | |
| Online Erişim: | https://doi.org/10.1016/j.nima.2017.08.054 |
| Materyal Türü: | Elektronik Kitap Bölümü |