Matching a line to an explosive-emission diode in local therapy devices; Technical Physics Letters; Vol. 28, iss. 7

Chi tiết về thư mục
Parent link:Technical Physics Letters
Vol. 28, iss. 7.— 2001.— [P. 550-551]
Tác giả chính: Mesyats G. A. Gennady Andreyevich
Tóm tắt:Title screen
It was previously suggested to use explosive-emission diodes generating X-ray pulses for local therapy purposes. The diode is usually powered by pulses supplied via a thin electrical cable. Effective operation of such a system requires that maximum pulsed energy transmitted via the cable would be absorbed in the diode. Taking into account nonlinearity of the current-voltage characteristic of the diode, it is impossible to ensure the line-to-diode matching in the general case. It is suggested to match the cable to the load by selecting a proper shape of the voltage pulse supplied to the diode. For an explosive-emission diode with flat electrodes, a virtually perfect matching is achieved with triangular voltage pulses.