Matching a line to an explosive-emission diode in local therapy devices

Podrobná bibliografie
Parent link:Technical Physics Letters
Vol. 28, iss. 7.— 2001.— [P. 550-551]
Hlavní autor: Mesyats G. A. Gennady Andreyevich
Shrnutí:Title screen
It was previously suggested to use explosive-emission diodes generating X-ray pulses for local therapy purposes. The diode is usually powered by pulses supplied via a thin electrical cable. Effective operation of such a system requires that maximum pulsed energy transmitted via the cable would be absorbed in the diode. Taking into account nonlinearity of the current-voltage characteristic of the diode, it is impossible to ensure the line-to-diode matching in the general case. It is suggested to match the cable to the load by selecting a proper shape of the voltage pulse supplied to the diode. For an explosive-emission diode with flat electrodes, a virtually perfect matching is achieved with triangular voltage pulses.
Режим доступа: по договору с организацией-держателем ресурса
Jazyk:angličtina
Vydáno: 2001
Témata:
On-line přístup:http://dx.doi.org/10.1134/1.1498781
Médium: Elektronický zdroj Kapitola
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=655678

Podobné jednotky