Summary:Title screen
The article describes the original algorithm for control the process of measuring thermo electromotive force. Proposed algorithm allows minimizing the effect of destabilizing factors at the manual testing of metals and alloys. It was proposed to take measurements at a time point when the supply voltage is zero, as well during the pause of PWM voltage of heating and at zero value of contact resistance between the electrodes and testing sample. The control algorithm provides sampling only valid data based on the above criteria, as well, data averaging and indication.