| Riassunto: | Title screen The paper presents the results on the study of the potassium nitrate dosimeter to estimate the characteristics of the pulsed electron radiation with an electron energy of 500 keV. One of the final products formed after potassium nitrate radiation is nitrite ion, whose radiation-chemical yield is constant and it equals 1.6 ion/100 eV. The measured concentration of nitrite ion with the Griess reagent enables to estimate the absorbed dose in any geometrical point of the irradiated volume. The comparable results of the magnitude of the absorbed dose obtained using a film dose color indicator are presented. The possibility to use an aqueous solution of potassium nitrate as a dosimeter, which is equivalent to the irradiated aqueous mediums, is studied. It is shown that the use of the potassium nitrate as a dosimeter is restricted due to the change in radiation-chemical yield of nitrite ions in solutions of different concentration. Режим доступа: по договору с организацией-держателем ресурса |
| Lingua: | inglese |
| Pubblicazione: |
2016
|
| Serie: | Applied engineering |
| Soggetti: | |
| Accesso online: | http://dx.doi.org/10.1109/IFOST.2016.7884322 |
| Natura: | Elettronico Capitolo di libro |
| KOHA link: | https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=651373 |
MARC
| LEADER | 00000nla2a2200000 4500 | ||
|---|---|---|---|
| 001 | 651373 | ||
| 005 | 20240111111542.0 | ||
| 035 | |a (RuTPU)RU\TPU\network\16622 | ||
| 035 | |a RU\TPU\network\16614 | ||
| 090 | |a 651373 | ||
| 100 | |a 20161110a2016 k y0rusy50 ba | ||
| 101 | 0 | |a eng | |
| 102 | |a RU | ||
| 105 | |a a z 101zy | ||
| 135 | |a drnn ---uucaa | ||
| 181 | 0 | |a i | |
| 182 | 0 | |a b | |
| 200 | 1 | |a Using the chemical dosimeter to measure the integral characteristics of the pulsed electron radiation |f L. R. Merinova, E. A. Proskurina, L. N. Shiyan | |
| 203 | |a Text |c electronic | ||
| 225 | 1 | |a Applied engineering | |
| 300 | |a Title screen | ||
| 320 | |a [References: p. 567 (4 tit.)] | ||
| 330 | |a The paper presents the results on the study of the potassium nitrate dosimeter to estimate the characteristics of the pulsed electron radiation with an electron energy of 500 keV. One of the final products formed after potassium nitrate radiation is nitrite ion, whose radiation-chemical yield is constant and it equals 1.6 ion/100 eV. The measured concentration of nitrite ion with the Griess reagent enables to estimate the absorbed dose in any geometrical point of the irradiated volume. The comparable results of the magnitude of the absorbed dose obtained using a film dose color indicator are presented. The possibility to use an aqueous solution of potassium nitrate as a dosimeter, which is equivalent to the irradiated aqueous mediums, is studied. It is shown that the use of the potassium nitrate as a dosimeter is restricted due to the change in radiation-chemical yield of nitrite ions in solutions of different concentration. | ||
| 333 | |a Режим доступа: по договору с организацией-держателем ресурса | ||
| 461 | 1 | |0 (RuTPU)RU\TPU\network\20157 |t 11th International Forum on Strategic Technology (IFOST 2016) |o 1-3 June 2016, Novosibirsk, Russiain |o in 2 pt. |o [proceedings] |f Novosibirsk State Technical University |d 2016 | |
| 463 | 0 | |0 (RuTPU)RU\TPU\network\20159 |t Pt. 2 |v [P. 564-567] |d 2016 | |
| 610 | 1 | |a электронный ресурс | |
| 610 | 1 | |a труды учёных ТПУ | |
| 610 | 1 | |a химические дозиметры | |
| 610 | 1 | |a импульсные электронные приборы | |
| 610 | 1 | |a нитрат калия | |
| 610 | 1 | |a водные растворы | |
| 610 | 1 | |a pulsed electron radiation | |
| 610 | 1 | |a dosimetry | |
| 610 | 1 | |a crystalline potassium nitrate dosimeter | |
| 610 | 1 | |a aqueous solution of potassium nitrate | |
| 610 | 1 | |a film dosimeter | |
| 700 | 1 | |a Merinova |b L. R. |c physicist |c Associate Scientist of Tomsk Polytechnic University |f 1981- |g Liliya Rashidovna |3 (RuTPU)RU\TPU\pers\32027 |9 16074 | |
| 701 | 1 | |a Proskurina |b E. A. |g Elena Aleksandrovna | |
| 701 | 1 | |a Shiyan |b L. N. |c chemist |c Associate Professor of Tomsk Polytechnic University, Candidate of physical and mathematical sciences |f 1952- |g Lyudmila Nikolaevna |3 (RuTPU)RU\TPU\pers\31523 | |
| 712 | 0 | 2 | |a Национальный исследовательский Томский политехнический университет (ТПУ) |b Институт физики высоких технологий (ИФВТ) |b Лаборатория № 1 |3 (RuTPU)RU\TPU\col\19035 |
| 712 | 0 | 2 | |a Национальный исследовательский Томский политехнический университет (ТПУ) |b Институт физики высоких технологий (ИФВТ) |b Кафедра общей химии и химической технологии (ОХХТ) |3 (RuTPU)RU\TPU\col\21253 |
| 801 | 2 | |a RU |b 63413507 |c 20171018 |g RCR | |
| 856 | 4 | |u http://dx.doi.org/10.1109/IFOST.2016.7884322 | |
| 942 | |c CF | ||