The effect of parameters of electrocardiograph measuring channel on recordings of micropotentials of the heart; Biology and Medicine; Vol. 6, iss. 2

Opis bibliograficzny
Parent link:Biology and Medicine
Vol. 6, iss. 2.— 2014.— [BM-025-14, 5 p.]
organizacja autorów: Национальный исследовательский Томский политехнический университет (ТПУ) Институт неразрушающего контроля (ИНК) Лаборатория № 63 (Медицинского приборостроения), Национальный исследовательский Томский политехнический университет (ТПУ) Институт неразрушающего контроля (ИНК) Кафедра физических методов и приборов контроля качества (ФМПК), Национальный исследовательский Томский политехнический университет (ТПУ) Институт кибернетики (ИК) Кафедра компьютерных измерительных систем и метрологии (КИСМ)
Kolejni autorzy: Avdeeva D. K. Diana Konstantinovna, Kazakov V. Yu. Veniamin Yurevich, Natalinova N. M. Nataliya Mikhailovna, Maksimov I. V. Ivan Vadimovich, Balahonova M. V. Marija Vjacheslavovna
Streszczenie:Title screen
The paper presents the research findings on the total effect of different types of filters on the electrocardiogram (ECG) signal recorded by a high-resolution electrocardiograph based on nanoelectrodes without filters in the band from 0 to 150 Hz. A standard block diagram of one ECG channel includes both amplifiers and a number of filters: a high-pass filter, a filter of constant component, a low-pass filter, an anti-aliasing filter, a rejection filter, and others. The method of computer simulation was used to study the effect of the fine structure distortions (low-amplitude high-frequency fluctuations) of the heart biopotentials caused by the filter circuits. The simulation was performed using the LabVIEW software package. It is shown that the filters distort the shape and amplitude of the heart micropotentials, and introduce phase shifts or remove them from the output signal of the considered ECG. It is essential to consider signal filtering when applying the amplitude and time diagnostic criteria. In some cases, application of diagnostic criteria to the filtered signal can produce erroneous results.
Język:angielski
Wydane: 2014
Hasła przedmiotowe:
Dostęp online:http://www.biolmedonline.com/Articles/Vol6_2_2014/BM-025-14_Avdeeva.pdf
Format: MixedMaterials Elektroniczne Rozdział
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=644317

MARC

LEADER 00000nla0a2200000 4500
001 644317
005 20250310163204.0
035 |a (RuTPU)RU\TPU\network\9382 
035 |a RU\TPU\network\7225 
090 |a 644317 
100 |a 20151110d2014 k||y0rusy50 ba 
101 0 |a eng 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a The effect of parameters of electrocardiograph measuring channel on recordings of micropotentials of the heart  |f D. K. Avdeeva [et al.] 
203 |a Text  |c electronic 
300 |a Title screen 
320 |a [References: 10 tit.] 
330 |a The paper presents the research findings on the total effect of different types of filters on the electrocardiogram (ECG) signal recorded by a high-resolution electrocardiograph based on nanoelectrodes without filters in the band from 0 to 150 Hz. A standard block diagram of one ECG channel includes both amplifiers and a number of filters: a high-pass filter, a filter of constant component, a low-pass filter, an anti-aliasing filter, a rejection filter, and others. The method of computer simulation was used to study the effect of the fine structure distortions (low-amplitude high-frequency fluctuations) of the heart biopotentials caused by the filter circuits. The simulation was performed using the LabVIEW software package. It is shown that the filters distort the shape and amplitude of the heart micropotentials, and introduce phase shifts or remove them from the output signal of the considered ECG. It is essential to consider signal filtering when applying the amplitude and time diagnostic criteria. In some cases, application of diagnostic criteria to the filtered signal can produce erroneous results. 
461 |t Biology and Medicine 
463 |t Vol. 6, iss. 2  |v [BM-025-14, 5 p.]  |d 2014 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
610 1 |a ЭКГ 
610 1 |a сердце 
610 1 |a фильтры 
701 1 |a Avdeeva  |b D. K.  |c specialist in the field of non-destructive testing  |c Professor of Tomsk Polytechnic University, Doctor of technical sciences  |f 1943-  |g Diana Konstantinovna  |3 (RuTPU)RU\TPU\pers\33062 
701 1 |a Kazakov  |b V. Yu.  |c specialist in the field of control and measurement equipment  |c Associate Professor of Tomsk Polytechnic University, Candidate of physical and mathematical sciences  |f 1959-  |g Veniamin Yurevich  |3 (RuTPU)RU\TPU\pers\33124  |9 16945 
701 1 |a Natalinova  |b N. M.  |c specialist in the field of Informatics and computer engineering  |c Associate Professor of Tomsk Polytechnic University, candidate of technical sciences  |f 1982-  |g Nataliya Mikhailovna  |3 (RuTPU)RU\TPU\pers\33817  |9 17411 
701 1 |a Maksimov  |b I. V.  |g Ivan Vadimovich 
701 1 |a Balahonova  |b M. V.  |g Marija Vjacheslavovna 
712 0 2 |a Национальный исследовательский Томский политехнический университет (ТПУ)  |b Институт неразрушающего контроля (ИНК)  |b Лаборатория № 63 (Медицинского приборостроения)  |3 (RuTPU)RU\TPU\col\19731 