Corrosion resistance of silicon-modified nitinol in artificial physiological solutions; Advanced Biomaterials and Devices in Medicine; Vol. 1, iss. 1

Podrobná bibliografie
Parent link:Advanced Biomaterials and Devices in Medicine
Vol. 1, iss. 1.— 2014.— [P. 46-52]
Korporativní autor: Национальный исследовательский Томский политехнический университет (ТПУ) Институт физики высоких технологий (ИФВТ) Кафедра общей и неорганической химии (ОНХ)
Další autoři: Lotkov A. I., Psakhie S. G. Sergey Grigorievich, Meisner L. L., Korshunov A. V. Andrey Vladimirovich, Meisner S. N., Diamant V. M.
Shrnutí:Title screen
Electrochemical methods were used to investigate the corrosion resistance of nitinol specimens modified with silicon by ion implantation (fluence 2  1017 ions/cm2). After ion beam treatment, nickel concentration in the surface layer is significantly reduced up to 20 nm deep and a silicon-containing layer is formed at the depth of 10...80 nm with the maximum concentration of 30 at % at the depth of 30...35 nm. The breakdown (pitting) potential Eb of such NiTi–Si specimens in 0.9% NaCl physiological solution and in artificial blood plasma is ~0.9 V (Ag/AgCl/ KCl sat.) which is much higher than Eb of test specimens subjected to mechanical, chemical or electrochemical treatment. It is shown that under potentiostatic conditions at Eb the NiTi–Si specimen surface is stable to failure and does not exhibit pitting, spotting or microcracking. The comparison of corrosion test results with the cyclic voltammetry data for NiTi, Ti and Ni as well as with thermodynamic calculations demonstrates that the surface of the modified NiTi–Si specimens has enhanced stability with respect to the release of nickel ions into solutions.
Jazyk:angličtina
Vydáno: 2014
Témata:
On-line přístup:http://abiodem.com/abstract.php?id=87
Médium: Elektronický zdroj Kapitola
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=642907

MARC

LEADER 00000nla0a2200000 4500
001 642907
005 20250211162605.0
035 |a (RuTPU)RU\TPU\network\7895 
090 |a 642907 
100 |a 20150821d2014 k||y0rusy50 ba 
101 0 |a eng 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a Corrosion resistance of silicon-modified nitinol in artificial physiological solutions  |f A. I. Lotkov [et al.] 
203 |a Text  |c electronic 
300 |a Title screen 
330 |a Electrochemical methods were used to investigate the corrosion resistance of nitinol specimens modified with silicon by ion implantation (fluence 2  1017 ions/cm2). After ion beam treatment, nickel concentration in the surface layer is significantly reduced up to 20 nm deep and a silicon-containing layer is formed at the depth of 10...80 nm with the maximum concentration of 30 at % at the depth of 30...35 nm. The breakdown (pitting) potential Eb of such NiTi–Si specimens in 0.9% NaCl physiological solution and in artificial blood plasma is ~0.9 V (Ag/AgCl/ KCl sat.) which is much higher than Eb of test specimens subjected to mechanical, chemical or electrochemical treatment. It is shown that under potentiostatic conditions at Eb the NiTi–Si specimen surface is stable to failure and does not exhibit pitting, spotting or microcracking. The comparison of corrosion test results with the cyclic voltammetry data for NiTi, Ti and Ni as well as with thermodynamic calculations demonstrates that the surface of the modified NiTi–Si specimens has enhanced stability with respect to the release of nickel ions into solutions. 
461 |t Advanced Biomaterials and Devices in Medicine 
463 |t Vol. 1, iss. 1  |v [P. 46-52]  |d 2014 
610 1 |a электронный ресурс 
610 1 |a труды учёных ТПУ 
701 1 |a Lotkov  |b A. I. 
701 1 |a Psakhie  |b S. G.  |c physicist  |c head of laboratory, Advisor to the rector, head of Department, Tomsk Polytechnic University, doctor of physico-mathematical Sciences  |f 1952-  |g Sergey Grigorievich  |3 (RuTPU)RU\TPU\pers\33038 
701 1 |a Meisner  |b L. L. 
701 1 |a Korshunov  |b A. V.  |c chemist  |c Professor of Tomsk Polytechnic University, Candidate of chemical sciences  |f 1970-  |g Andrey Vladimirovich  |3 (RuTPU)RU\TPU\pers\33141 