Digital Signal Processing in Pulse Method for Measuring Frequency Response Function of High-Current Shunt; Advances in Instrumentation and Sensors Interoperability

التفاصيل البيبلوغرافية
Parent link:Advances in Instrumentation and Sensors Interoperability.— 2013.— [P. 496-500]
المؤلف الرئيسي: Zarevich A. I. Anton Ivanovich
مؤلف مشترك: Национальный исследовательский Томский политехнический университет (ТПУ) Институт кибернетики (ИК) Кафедра компьютерных измерительных систем и метрологии (КИСМ)
مؤلفون آخرون: Muravyov (Murav’ev) S. V. Sergey Vasilyevich, Bedareva E. V. Elena Vyacheslavovna
الملخص:Title screen
In the paper a digital signal processing method for measuring frequency response function of highcurrent shunt are described. This is based on a joint digital processing of the input short pulse signal and a corresponding output. In order to measure the frequency response function, a short current pulse is applied to the shunt input; input and output signals of the shunt are recorded in a digital oscilloscope memory; then spectra of these two signals are computed. Amplitude and phase responses are determined by the arithmetic ratio of spectral components of the input and output signals at appropriate frequencies. For this purpose, an algorithm has been proposed and investigated.
اللغة:الإنجليزية
منشور في: 2013
الموضوعات:
الوصول للمادة أونلاين:http://www.imeko.org/publications/tc4-2013/IMEKO-TC4-2013-120.pdf
التنسيق: الكتروني فصل الكتاب
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=641939