The research of characteristics of multichannel GaAs detectors; 2011 International Siberian Conference on Control and Communications (SIBCON), Russia, Krasnoyarsk, September 15−16, 2011

Detalhes bibliográficos
Parent link:2011 International Siberian Conference on Control and Communications (SIBCON), Russia, Krasnoyarsk, September 15−16, 2011.— 2011.— [P. 247 - 248]
Outros Autores: Nam I. F. Irina Feliksovna, Rjabkov S. A., Tolbanov O. P., Tyazhev A. V.
Resumo:Title screen
In this paper the results of multichannel GaAs detectors testing in integral mode are presented. It is shown that for x-ray quanta at the energy range 40-75 keV and exposure dose more than 1,5 mR the noise characteristics of the detectors are mostly depends on photon noise. It is experimentally proved that negative potential on irradiated contact and an inclination of detectors at a small angle to a direction of falling of radiation allow to increase sensitivity of detectors considerably.
Режим доступа: по договору с организацией-держателем ресурса
Idioma:inglês
Publicado em: 2011
Assuntos:
Acesso em linha:http://dx.doi.org/10.1109/SIBCON.2011.6072645
Formato: Recurso Electrónico Capítulo de Livro
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=639939