A cryogenic calorimeter based on a superconducting phase transition thermometer with thermal feedback and SQUID read out; Superconductor Science and Technology; Vol. 12, iss. 11

書誌詳細
Parent link:Superconductor Science and Technology.— , 1988-
Vol. 12, iss. 11.— 1999.— [P. 1033-1035]
その他の著者: Meier O., Probst F., Seidel W., Schnagl J., Sisti M., Uchaykin S. V. Sergey Victorovich
要約:Title screen
Cryogenic calorimeters with superconducting phase transition (SPT) thermometers are extremely sensitive particle detectors. Some characteristics of SPT calorimeters such as linearity and dynamic range often suffer from the superconducting transition shape. The common way to improve the performances is to use the voltage bias to operate the SPT calorimeter in strong electrothermal feedback mode. This method does not allow us to optimize the noise, linearity and dynamic properties simultaneously. We have developed a novel SQUID circuit with thermal feedback which allows to improve the linearity, dynamic range and speed of the SPT calorimeter. Results obtained with a cryogenic calorimeter based on a tungsten thermometer are presented
Режим доступа: по договору с организацией-держателем ресурса
言語:英語
出版事項: 1999
主題:
オンライン・アクセス:http://elibrary.ru/item.asp?id=13321630
http://iopscience.iop.org/0953-2048/12/11/402