We have developed massive cryogenic particle detectors to be used in the CRESST dark matter search. Each detector is made of a sapphire crystal and a tungsten superconducting phase transition thermometer. In this paper, we report on the results obtained with four 262 g detectors, which show energy thresholds as low as 350 e V and good energy resolution at low energies. The shape of the experimental pulses, the linearity of the detector response and the energy dependence of the resolution are discussed
Режим доступа: по договору с организацией-держателем ресурса
Language:English
Published: 2002
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S0168900201008014
http://elibrary.ru/item.asp?id=14121138
Format: Electronic Book Chapter
KOHA link:https://koha.lib.tpu.ru/cgi-bin/koha/opac-detail.pl?biblionumber=637180

MARC

LEADER 00000nla0a2200000 4500
001 637180
005 20250415164705.0
035 |a (RuTPU)RU\TPU\network\1277 
035 |a RU\TPU\network\1261 
090 |a 637180 
100 |a 20140602d2001 k||y0rusy50 ba 
101 0 |a eng 
102 |a NL 
135 |a drcn ---uucaa 
181 0 |a i  
182 0 |a b 
200 1 |a Massive cryogenic particle detectors with low energy threshold  |f M. Sisti [et al.] 
203 |a Text  |c electronic 
300 |a Title screen 
330 |a We have developed massive cryogenic particle detectors to be used in the CRESST dark matter search. Each detector is made of a sapphire crystal and a tungsten superconducting phase transition thermometer. In this paper, we report on the results obtained with four 262 g detectors, which show energy thresholds as low as 350 e V and good energy resolution at low energies. The shape of the experimental pulses, the linearity of the detector response and the energy dependence of the resolution are discussed 
333 |a Режим доступа: по договору с организацией-держателем ресурса 
461 |t Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment  |o Scientific Journal 
463 |t Vol. 466, iss. 3  |v [P. 499-508]  |d 2002 
610 1 |a труды учёных ТПУ 
610 1 |a электронный ресурс 
701 1 |a Sisti  |b M. 
701 1 |a Meier  |b O. 
701 1 |a Buhler  |b M. 
701 1 |a Cooper  |b S. 
701 1 |a Jorgens  |b V. 
701 1 |a Loidl  |b M. 
701 1 |a Nagel  |b U. 
701 1 |a Probst  |b F. 
701 1 |a Seidel  |b W. 
701 1 |a Stolovits  |b A. 
701 1 |a Stodolsky  |b L. 
701 1 |a Uchaykin  |b S. V.  |c specialist in the field of non-destructive testing  |c Engineer of Tomsk Polytechnic University, Doctor of physical and mathematical sciences  |f 1963-  |g Sergey Victorovich  |3 (RuTPU)RU\TPU\pers\32279 
701 1 |a Zerle  |b L. 
801 2 |a RU  |b 63413507  |c 20160512  |g RCR 
850 |a 63413507 
856 4 |u http://www.sciencedirect.com/science/article/pii/S0168900201008014 
856 4 |u http://elibrary.ru/item.asp?id=14121138 
942 |c CF